Quantidex qPCR BCR-ABL IS Kit
Bcr/Abl1 Monitoring Test. De Novo DEN160003, from Asuragen, Inc. (Austin, TX). Product code OYX, class 2.
As of Nov 27, 2026, DEN160003 is an FDA De Novo held by Asuragen, Inc., granted on Jul 22, 2016 after 185 days, with no recalls tied to it in FDA's recall database.
- Review time: 185 days
- Recalls: 0
- Cited as a predicate by: 2 devices
- Design
- Retrospective, multi-center clinical outcome study of CML patients with EFS by MR3.0 threshold at 36 months
- Enrolled
- 98
- Where
- 2 sites, US
- Primary endpoint
- Difference in 36-month event-free survival probability between patients attaining MR3.0 and not attaining MR3.0; success requires estimate >10 percentage points and significance at one-sided 2.5% level
- Result
- Met: EFS difference 22.2% (95% CI 2.0%-42.4%), p=0.0279
- Xpert BCR-ABL Ultra, GeneXpert Dx System, GeneXpert Infinity-48s and GeneXpert Infinity-80 SystemsK190076 · Cepheid · Sep 27, 2019
- QXDx BCR-ABL %IS Kit for use on the QXDx AutoDG ddPCR SystemK181661 · Bio-Rad Laboratories, Inc. · Feb 13, 2019
- Fastplex™ BCR-ABL (p210) %IS digital PCR KitK252243 · Suzhou Precigenome, Ltd. Co; · Jul 2, 2026
- BCR-ABL1 (p210) % IS Kit (Digital PCR Method)K221869 · Suzhou Sniper Medical Technologies Co., Ltd. · Sep 5, 2023
- Type
- Direct
- Received
- Jan 19, 2016
- Review panel
- Medical Genetics
- Regulation
- 21 CFR 866.6060
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files