Modified Trevo ProVue Retriever, Trevo XP ProVue Retriever 4x20mm, Trevo XP ProVue Retriever 3x20mm, Trevo XP ProVue Retriever 6x25mm, Trevo XP ProVue Retriever 4x30mm

Neurovascular Mechanical Thrombectomy Device For Acute Ischemic Stroke Treatment. De Novo DEN150049, from Concentric Medical, Inc. (Mountain View, CA). Product code POL, class 2.

As of Nov 19, 2026, DEN150049 is an FDA De Novo held by Concentric Medical, Inc., granted on Sep 2, 2016 after 312 days, with no recalls tied to it in FDA's recall database.

  • Review time: 312 days
  • Recalls: 0
  • Cited as a predicate by: 2 devices

Pivotal study

The main clinical evidence FDA relied on, as its decision summary describes it
Design
Prospective, randomized, open label, controlled, multicenter trial (MR CLEAN) of thrombectomy plus medical management vs. medical management alone
Enrolled
500
Where
16 sites, Netherlands
Primary endpoint
Functional independence (mRS ≤2) at 90 days for Trevo thrombectomy plus MM vs. MM alone; primary safety: all-cause mortality at 90 days
Result
Trevo 28/96 (29.2%) vs. MM 48/249 (19.3%) with mRS ≤2 at 90 days; OR 1.88 [1.07, 3.29], p=0.014. Mortality 13.3% (16/120) Trevo vs. 22.89% (57/249) MM

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Other POL clearances

The newest 510(k)s for the same product code; their median review was 153 days; see every POL device

Elsewhere

Record

Type
Direct
Received
Oct 26, 2015
Review panel
Neurology
Regulation
21 CFR 882.5600

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files