WallFlex Biliary RX Fully Covered Stent System RMV
Biliary Stent System For Benign Strictures. De Novo DEN150040, from Boston Scientific (Marlboro, MA). Product code PNB, class 2.
As of Nov 20, 2026, DEN150040 is an FDA De Novo held by Boston Scientific, granted on Jun 3, 2016 after 280 days, with no recalls tied to it in FDA's recall database.
- Review time: 280 days
- Recalls: 0
- Cited as a predicate by: 1 device
- Design
- Prospective, nonrandomized, multinational study; chronic pancreatitis cohort (NCT01014390)
- Enrolled
- 127
- Where
- 13 sites, 11 countries outside the US
- Primary endpoint
- Stent removability: endoscopic removal without serious stent removal-related adverse events, assessed to 1 month post-removal
- Result
- Stent removability 72.4% (92/127) ITT and 78.0% (92/118) PP; no performance goal stated in text
- WallFlex Biliary RX Stent System; Epic Biliary Stent System; WallFlex Biliary RX Fully Covered Stent System RMVK240464 · Boston Scientific · Aug 30, 2024
- Type
- Direct
- Received
- Aug 28, 2015
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.5011
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files