Heartflow Ffrct
Coronary Vascular Physiologic Simulation Software. De Novo DEN130045, from Heartflow (Redwood City, CA). Product code PJA, class 2.
As of Nov 20, 2026, DEN130045 is an FDA De Novo held by Heartflow, granted on Nov 26, 2014 after 387 days, with no recalls tied to it in FDA's recall database.
- Review time: 387 days
- Recalls: 0
- Cited as a predicate by: 2 devices
- Design
- Prospective, multicenter, non-randomized study (HFNXT) vs. invasive FFR reference standard
- Enrolled
- 276
- Where
- 11 sites, 8 countries in Canada, Europe and Asia
- Primary endpoint
- Per-vessel sensitivity and specificity of FFRCT vs. invasive FFR (<=0.80) reference standard; pre-specified targets 65% and 55%
- Result
- Met: per-vessel sensitivity 83.5% (lower 95% CI 75.3%) vs. 65% goal; specificity 85.8% (lower 95% CI 81.5%) vs. 55% goal
- Elucid FFR-IQ (B.1)K262015 · Elucid Bioimaging, Inc. · Sep 29, 2026
- FFRctK152733 · HeartFlow, Inc. · Jan 13, 2016
- Elucid FFR-IQ (B.1)K262015 · Elucid Bioimaging, Inc. · Sep 29, 2026
- HeartFlow AnalysisK250902 · HeartFlow, Inc. · Jul 18, 2025
- HeartFlow AnalysisK213857 · HeartFlow, Inc. · Oct 14, 2022
- Deepvessel FfrK213657 · Keyamed Na, Inc. · Apr 1, 2022
- HeartFlow AnalysisK203329 · HeartFlow, Inc. · Jan 8, 2021
- Document
- De Novo decision summary (PDF) ↗
- Type
- Direct
- Received
- Nov 4, 2013
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1415
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files