Virulite Cold Sore Machine
Light Based Treatment For Cold Sores Herpes Simplex Virus-1. De Novo DEN090012, from Virulite, LLC (Irvine, CA). Product code OKJ, class 2.
As of Nov 25, 2026, DEN090012 is an FDA De Novo held by Virulite, LLC, granted on Oct 18, 2012 after 1,206 days, with no recalls tied to it in FDA's recall database.
- Review time: 1,206 days
- Recalls: 0
- Cited as a predicate by: 1 device
- Design
- Double-blind, placebo-controlled, randomized clinical trial of ViruLite vs. placebo (subjects 20 years and older)
- Enrolled
- 95
- Primary endpoint
- Time to healing (subject-described re-epithelialization) from onset of cold sore; safety assessed for redness, discomfort, burns, blistering
- Result
- Randomized dataset (95 randomized; 48 active) did not by itself demonstrate reasonable assurance of effectiveness due to data concerns; no numeric healing-time results stated
- Cold Sore Device (Model: QPZ-01)K222205 · Light Tree Ventures Europe B.V. · Oct 7, 2022
- Luminance Red Cold Sore Device (TN1927G)K251973 · Luminance Medical Ventures, Inc. · Sep 23, 2025
- Cold Sore Device (QPZ-03)K241155 · Shenzhen Nuon Medical Equipment Co., Ltd. · Sep 5, 2024
- Cold Sore Device (Model: QPZ-01)K222205 · Light Tree Ventures Europe B.V. · Oct 7, 2022
- Type
- Post-NSE
- Received
- Jun 30, 2009
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.4860
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files