FDA establishment 3014133165, Holliston, Massachusetts, United States. Part of NeuroSync, Inc.'s FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 2
Official action indicated: 1
Last inspection ended: Feb 5, 2025
Warning letters: 2
Inspections by Fiscal Year
FY2018 to FY2027 to date; 1 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 801.20(a). Label to bear a unique device identifier The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
21 CFR 820.30(c). Design input - Lack of or inadequate procedures Procedures for design input have not been adequately established.
21 CFR 820.30(i). Design changes - Lack of or Inadequate Procedures Procedures for design change have not been adequately established.
21 CFR 820.80(a). Lack of or inadequate procedures - Acceptance activities Procedures for acceptance activities have not been adequately established.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 830.300(a). Devices subject to device identification GUDID data submission requirements. The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
21 CFR 806.20(b)(4). Justification for not reporting A justification for not reporting the correction or removal action to FDA that included conclusions by a designated person was not included in the record.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.30(g). Design Validation - Risk analysis not performed/inadequate Risk analysis is inadequate.
21 CFR 820.184. Lack of or inadequate DHR procedures Procedures for device history records have not been adequately established.
21 CFR 820.30(f). Design verification - output does not meet input requirement Design verification does not confirm that design output meets design input requirements.
21 CFR 820.100(b). Documentation Corrective and preventive action activities and/or results have not been adequately documented.
21 CFR 820.198(c). Investigation of device failures Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
21 CFR 820.30(d). Design output - documentation Design output was not adequately documented before release.
21 CFR 820.50(a)(1). Documented evaluation The evaluation of potential suppliers, contractors and consultants was not documented.
21 CFR 820.30(e). Design review - Lack of or inadequate procedures Procedures for design review have not been adequately established.
No Action Indicated (NAI)
Warning Letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm.FDA Data Dashboard and warning letters