FDA establishment 3012421755, Henrietta, New York, United States. Part of Turbett Surgical's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 4
Official action indicated: 0
Last inspection ended: Nov 6, 2025
Warning letters: 0
Inspections by Fiscal Year
FY2018 to FY2027 to date; 0 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.90(b)(2). Product rework procedures, Lack of or inadequate procedures Procedures for rework of nonconforming product have not been established.
21 CFR 820.100(b). Documentation Corrective and preventive action activities and/or results have not been adequately documented.
21 CFR 820.250(a). Statistical techniques - Lack of or inadequate procedures Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.20(c). Management review - Lack of or inadequate procedures Procedures for management review have not been adequately established.
21 CFR 820.25(b). Training records Personnel training is not documented.
21 CFR 820.30(g). Design validation- Lack of or inadequate procedures Procedures for design validation have not been adequately established.
21 CFR 820.30(i). Design changes - Lack of or Inadequate Procedures Procedures for design change have not been adequately established.
21 CFR 820.75(a). Lack of or inadequate process validation A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.22. Quality audits - Lack of or inadequate procedures Procedures for quality audits have not been adequately established.
21 CFR 820.30(h). Design transfer - Lack of or inadequate procedures Procedures for design transfer have not been adequately established.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.30(g). Design validation- Lack of or inadequate procedures Procedures for design validation have not been adequately established.
21 CFR 806.10(b). Time to report - 10 days A report of the required information regarding device correction and removal actions was not sent to FDA within 10 days of initiating the correction or removal.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.30(i). Design changes - Lack of or Inadequate Procedures Procedures for design change have not been adequately established.
21 CFR 803.50(a)(2). Individual Report of Malfunction An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
21 CFR 820.75(a). Lack of or inadequate process validation A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm.FDA Data Dashboard and warning letters