Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.
FDA establishment 3006111110, Albuquerque, New Mexico, United States. Part of Meditherm, Inc.'s FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 2
Official action indicated: 1
Last inspection ended: Sep 2, 2022
Warning letters: 0
Inspections by Fiscal Year
FY2018 to FY2027 to date; 1 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.30(i). Design changes - Lack of or Inadequate Procedures Procedures for design change have not been established.
21 CFR 820.80(b). Lack of or inadequate receiving acceptance procedures Procedures for acceptance of incoming product have not been established.
21 CFR 820.22. Quality Audit/Reaudit - conducted Quality audits have not been performed.
21 CFR 820.20(c). Management review - defined interval, sufficient frequency Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system with sufficient frequency.
21 CFR 820.50(a). Evaluation of suppliers, contractors, etc., requirements Requirements that must be met by suppliers have not been established.
21 CFR 820.180(b). Retention period Required records are not retained for the design and expected life of the device.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been established to control product that does not conform to specified requirements.
21 CFR 820.181. DMR - not or inadequately maintained A device master record has not been maintained.
21 CFR 803.17. Lack of Written MDR Procedures Written MDR procedures have not been developed.
21 CFR 820.198(e). Records of complaint investigation Records of complaint investigations do not include required information.
21 CFR 820.20(a). Quality policy and objectives The quality policy and quality objectives was/were not established by management with executive responsibility.
21 CFR 820.30(j). Design history file The design history file was not established and does not demonstrate that the design was developed following the requirements of 21 CFR 820.
21 CFR 820.30(a). Design control - no procedures Procedures for design control have not been established.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
21 CFR 820.40. Lack of procedures, or not maintained Document control procedures have not been established.
21 CFR 820.20(c). Management review - Lack of or inadequate procedures Procedures for management review have not been established.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
21 CFR 820.80(d). Lack of or inadequate final acceptance procedures Procedures for finished device acceptance have not been established.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been established.
21 CFR 820.22. Quality audits - Lack of or inadequate procedures Procedures for quality audits have not been established.
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm.FDA Data Dashboard, warning letters and enforcement reports