Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.
FDA establishment 2080783, Irvine, California, United States. Part of Nihon Kohden America's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 5
Official action indicated: 0
Last inspection ended: Oct 4, 2024
Warning letters: 0
Inspections by Fiscal Year
FY2018 to FY2027 to date; 0 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.50(a)(2). Supplier oversight The type and extent of control to be exercised over the product was not clearly defined.
21 CFR 820.198(c). Investigation of device failures Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
21 CFR 803.40(a). Report of death, injury The importer failed to submit a report to FDA on FDA form 3500A, with a copy to the manufacturer, within 30 days after receiving information that one of its marketed devices may have caused or contributed to a serious injury.
21 CFR 820.25(b). Training - Lack of or inadequate procedures Procedures for training and identifying training needs have not been adequately established.
21 CFR 820.100(b). Documentation Corrective and preventive action activities and/or results have not been adequately documented.
21 CFR 803.50(b)(1). Reporting Information Reasonably Known An MDR report submitted to FDA did not include all information that was reasonably known to the manufacturer.
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm.FDA Data Dashboard and warning letters