Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.
FDA establishment 1824226, Noblesville, Indiana, United States. Part of King Systems Corp.'s FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 8
Official action indicated: 2
Last inspection ended: May 16, 2023
Warning letters: 2
Inspections by Fiscal Year
FY2018 to FY2027 to date; 1 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 806.10(a)(1). Report of risk to health A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
21 CFR 820.250(b). Sampling plans Sampling plans are not written and based on valid statistical rationale.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.75(a). Lack of or inadequate process validation A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
21 CFR 820.80(c). Lack of or inadequate In-process acceptance procedures Procedures for the acceptance of in-process product have not been adequately established.
21 CFR 806.10(a)(2). Report of violation of the Act (see 803.52(e)(9)) A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.
21 CFR 820.75(a). Lack of or inadequate process validation A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Official Action Indicated (OAI)
Oct 29, 2012
Devices · Compliance: Devices
3 cited references
21 CFR 820.80(b). Lack of or inadequate receiving acceptance procedures Procedures for acceptance of incoming product have not been adequately established.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been adequately established to control product that does not conform to specified requirements.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 803.50(a)(2). Individual Report of Malfunction An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Voluntary Action Indicated (VAI)
Warning Letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm.FDA Data Dashboard and warning letters