Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.
FDA establishment 1044692, Casselberry, Florida, United States. Part of Magnatone Hearing Aid Corp.'s FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 4
Official action indicated: 0
Last inspection ended: Jul 14, 2025
Warning letters: 0
Inspections by Fiscal Year
FY2018 to FY2027 to date; 0 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.22. Quality Audit/Reaudit - conducted Quality audits have not been performed.
21 CFR 803.17(a)(2). Lack of System for Determining MDR Events The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
21 CFR 820.20(a). Management ensuring quality policy is understood Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.40(b). Change records, content Records of changes did not include a description of the change.
21 CFR 820.30(g). Design validation - Risk analysis Results of the design risk analysis were not adequately documented.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
21 CFR 820.30(c). Design input - Lack of or inadequate procedures Procedures for design input have not been adequately established.
21 CFR 820.30(i). Design changes - Lack of or Inadequate Procedures Procedures for design change have not been adequately established.
21 CFR 820.30(g). Design validation- Lack of or inadequate procedures Procedures for design validation have not been adequately established.
21 CFR 820.70(a). Production processes Production processes were not developed, controlled and monitored to ensure that a device conforms to its specifications.
21 CFR 803.17(a)(2). Lack of System for Determining MDR Events The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been established.
21 CFR 820.80(d). Lack of or inadequate final acceptance procedures Procedures for finished device acceptance have not been established.
21 CFR 803.17. Lack of Written MDR Procedures Written MDR procedures have not been implemented.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been established.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm.FDA Data Dashboard, warning letters and enforcement reports