FDA establishment 1000305979, Oregon, Ohio, United States. Part of R-Med, Inc.'s FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 2
Official action indicated: 0
Last inspection ended: Apr 23, 2025
Warning letters: 0
Inspections by Fiscal Year
FY2018 to FY2027 to date; 0 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.20(c). Management review - defined interval, sufficient frequency Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system.
21 CFR 820.30(g). Design Validation - Risk analysis not performed/inadequate Risk analysis is inadequate.
21 CFR 820.184. DHR - not or inadequately maintained A device history record has not been adequately maintained.
21 CFR 820.181. DMR - not or inadequately maintained A device master record has not been adequately maintained.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been adequately established to control product that does not conform to specified requirements.
21 CFR 820.198(a). Complaints Complaint files are not adequately maintained.
21 CFR 820.22. Quality Audit/Reaudit - conducted Quality audits have not been performed.
21 CFR 820.250(a). Statistical techniques - Lack of or inadequate procedures Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.
21 CFR 820.30(j). Design history file The design history file was not established.
21 CFR 820.80(a). Lack of or inadequate procedures - Acceptance activities Procedures for acceptance activities have not been established.
21 CFR 820.184. DHR - not or inadequately maintained A device history record has not been adequately maintained.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been adequately established to control product that does not conform to specified requirements.
21 CFR 820.198(a). Complaints Complaint files are not adequately maintained.
21 CFR 820.30(j). Design history file The design history file was not established.
21 CFR 820.250(a). Statistical techniques - Lack of or inadequate procedures Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.
21 CFR 820.181. DMR - not or inadequately maintained A device master record has not been maintained.
21 CFR 820.30(i). Design changes - Lack of or Inadequate Procedures Procedures for design change have not been established.
21 CFR 820.22. Quality Audit/Reaudit - conducted Quality audits have not been performed.
21 CFR 820.80(b). Lack of or inadequate receiving acceptance procedures Procedures for acceptance of incoming product have not been established.
21 CFR 820.20(c). Management review - defined interval, sufficient frequency Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system.
21 CFR 820.30(g). Design Validation - Risk analysis not performed/inadequate Risk analysis was not performed.
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm.FDA Data Dashboard and warning letters