 "method": [
  {
   "value": "34,279",
   "label": "510(k), De Novo and PMA summaries decided since 2016, read in full"
  },
  {
   "value": "412",
   "label": "cite a named modeling technique in a testing sentence"
  },
  {
   "value": "391 of 412",
   "label": "evidence sentences found word for word in FDA's documents; only these are shown"
  }
 ],
 "hint": "Open a period to see the summaries behind its count. The highlight marks the modeling technique the sentence names.",
 "rows": [
  {
   "id": "2016-2018",
   "label": "2016 to 2018: 1.6% of summaries",
   "count": 147,
   "records": [
    {
     "title": "CoreLink F3D™ Lateral System",
     "meta": [
      "Corelink",
      "19 December 2018",
      "K183239",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance Testing: FEA was performed to demonstrate that the modified geometries of the cage did not change the worst case device in the system.",
       "mark": [
        21,
        24
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K183239",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "FFRangio System",
     "meta": [
      "Cathworks",
      "19 December 2018",
      "K182149",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "A series of in vitro and animal studies were conducted by CathW orks to validate the computational modelling methods for the FFRangio system.",
       "mark": [
        85,
        108
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K182149",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Avenir Complete Hip System",
     "meta": [
      "Zimmer",
      "7 December 2018",
      "K182048",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Finite Element Analysis for Fatigue Testing 2.",
       "mark": [
        0,
        14
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K182048",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Renovis Tesera C/Tesera SC Anterior Cervical Fusion (ACF) System",
     "meta": [
      "Renovis Surgical Technologies",
      "7 December 2018",
      "K182007",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Differences between the Renovis Tesera C/Tesera SC ACF System and the predicate devices were assessed under Design Control by risk analysis and FEA raises no new safety or effectiveness questions when compared to the predicate device.",
       "mark": [
        144,
        147
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K182007",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Incraft(r) AAA Stent Graft System",
     "meta": [
      "Cordis US",
      "27 November 2018",
      "P150002",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "To support the root cause analysis, a range of axial compression values was determined to use as boundary conditions to input into a finite element analysis (FEA).",
       "mark": [
        133,
        147
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P150002",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Halifax Imaging Kit",
     "meta": [
      "Halifax Biomedical",
      "31 October 2018",
      "K182880",
      "Radiology"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Finite element analysis was performed on the support structures followed by design verification reviews.",
       "mark": [
        0,
        14
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K182880",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Valiant Navion Thoracic Stent Graft System",
     "meta": [
      "Medtronic Vascular",
      "19 October 2018",
      "P100040S036",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "To calculate fatigue safety The safety factors based on Finite Element factors by quantifying the the endurance limit shall be Pass Analysis strain on the stent when > 1. subject to in vivo conditions. 1.",
       "mark": [
        56,
        70
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P100040S036",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Eluvia Drug-Eluting Vascular Stent System",
     "meta": [
      "Boston Scientific",
      "18 September 2018",
      "P180011",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "The FEA analysis fatigue analysis with a safety factor > Analysis) simulated physiological 1. conditions in the SFA.",
       "mark": [
        4,
        7
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P180011",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Sypher Spacer System",
     "meta": [
      "Biogennix",
      "18 September 2018",
      "K181337",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "All devices compare similarly in: • Design features • Intended use • Materials • Dimensions • Function Mechanical Testing: Substantial Equivalence is supported by the results of mechanical testing and Finite Element Analysis (FEA).",
       "mark": [
        201,
        215
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K181337",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "LnK Lumbar Interbody Fusion Cage System",
     "meta": [
      "L&K BIOMED",
      "11 September 2018",
      "K181380",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "We performed finite element analysis to demonstrate that the additional components do not introduce a new worst case in terms of mechanical performance.",
       "mark": [
        13,
        27
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K181380",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "NuVasive Brigade Lateral System",
     "meta": [
      "Nu Vasive, Incorporated",
      "24 August 2018",
      "K181386",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance Data Finite Elemental Analysis (FEA) and engineering rationale were provided as evidence that the modified design of the Brigade® Lateral System construct did not create a new worst case for performance testing.",
       "mark": [
        17,
        31
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K181386",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Sonata Sonography-Guided Transcervical Fibroid Ablation System",
     "meta": [
      "Gynesonics",
      "15 August 2018",
      "K173703",
      "Obstetrics/Gynecology"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Development Effect of Parameter Variations on Ablation Dimensions - and Validation Variation in Tissue Parameters Variation in Treatment Temperature Variation in RF Power Variation in Treatment Duration Finite Element Method (FEM) Computational Modeling to determine sensitivity of ablation dimensions to variations in tissue parameters.",
       "mark": [
        203,
        217
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K173703",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "MotoCLIP/HiMAX Step Staple Implant System",
     "meta": [
      "Crossroads Extremity Systems",
      "9 August 2018",
      "K181866",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance Testing: Finite element analysis was performed utilizing the worst-case sizes the subject MotoCLIP™/HiMAX™ Step Staple Implant System.",
       "mark": [
        21,
        35
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K181866",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "EVOS Wrist Fracture Plating System",
     "meta": [
      "Smith & Nephew",
      "9 August 2018",
      "K181533",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Summary of Pre-clinical Testing x\u0003 Finite element analysis (FEA) was conducted on the proposed plate designs to determine the worst case plates for further mechanical testing.",
       "mark": [
        35,
        49
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K181533",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "MyPlant II Implant System",
     "meta": [
      "Hager& Meisinger",
      "9 August 2018",
      "K173819",
      "Dental"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "The effects of this design change have been evaluated by fatigue testing and FEM analysis, supporting substantial equivalence.",
       "mark": [
        77,
        80
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K173819",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "MotoCLIP/HiMAX Implant System",
     "meta": [
      "Crossroads Extremity Systems",
      "29 June 2018",
      "K181410",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance Testing: Theoretical analysis (FEA) of the worst case MotoCLIP™ /HiMAX™ Implant System was performed to predict compression force as well as max imum strain for the subject and predicate devices.",
       "mark": [
        43,
        46
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K181410",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Juliet Ti",
     "meta": [
      "Spineart",
      "28 June 2018",
      "K181203",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "A Finite Element Analysis has been submitted to support substantial equivalence.",
       "mark": [
        2,
        16
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K181203",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "iStent inject Trabecular Micro-Bypass System (Model G2-M-IS)",
     "meta": [
      "Glaukos",
      "21 June 2018",
      "P170043",
      "Ophthalmic"
     ],
     "tags": [
      {
       "label": "Computational fluid dynamics"
      },
      {
       "label": "Finite element analysis"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Pressure/Flow Stent has negligible flow resistance Pass Numerical modeling, Characteristics including computational fluid dynamics, was used to evaluate the flow through the stents over physiologically relevant boundary conditions.",
       "mark": [
        56,
        74
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P170043",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "SALVATION External Fixation System",
     "meta": [
      "Wrightmedicaltechnologyinc",
      "6 June 2018",
      "K180832",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Substantial Equivalence – Non-Clinical Evidence Mechanical Pull-out testing, Finite Element Analysis, and Engineering analysis demonstrated substantial equivalence comparing each modified component to the predicate components.",
       "mark": [
        77,
        91
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K180832",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Low-Profile Visualized Intraluminal Support (LVIS) and LVIS Jr.",
     "meta": [
      "MicroVention",
      "30 May 2018",
      "P170013",
      "Neurology"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "The finite element analysis (FEA) predicted the LVIS/LVIS Jr. fatigue lifetime has adequate safety built into the design with a safety factor of > 2.39.",
       "mark": [
        4,
        18
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P170013",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "EVEREST Spinal System",
     "meta": [
      "K2m",
      "30 May 2018",
      "K181188",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Primary Predicate: EVEREST Spinal System (K2M Inc. – K161369) Performance Data: The modified device was evaluated via finite element analyses.",
       "mark": [
        118,
        132
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K181188",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Reliance Lumber IBF System",
     "meta": [
      "Reliance Medical Systems",
      "15 May 2018",
      "K180687",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Finite Element Analysis has been carried out to access the effect of new material on the structural properties of the implant.",
       "mark": [
        0,
        14
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K180687",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Pro-Link® Ti Stand-Alone Cervical Spacer System",
     "meta": [
      "Life Spine",
      "10 May 2018",
      "K180642",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance Data: Engineering rationales and FEA were provided to demonstrate the substantial equivalency of the Life Spine PRO-LINK Ti Stand-Alone Cervical Spacer System.",
       "mark": [
        45,
        48
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K180642",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "A-CIFT SoloFuse",
     "meta": [
      "Spinefrontier",
      "8 May 2018",
      "K172484",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Failure Effects Analysis (FEA) was used in lieu of mechanical testing to show substantial equivalence to the previously cleared design.",
       "mark": [
        26,
        29
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K172484",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Arthrex Shoulder System",
     "meta": [
      "Arthrex",
      "29 March 2018",
      "K171841",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "MRI RF and gradient heating simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance Data Non-clinical testing and in-vivo electromagnetic simulation per the FDA Guidance, “Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment”, demonstrated that the Arthrex Shoulder Systems are “MR Conditional”.",
       "mark": [
        50,
        76
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K171841",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Memo 4D",
     "meta": [
      "Sorin Group Italia",
      "16 March 2018",
      "K180411",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Non-clinical testing included: - computational stress analysis (FEA) and tensile strength testing; - fatigue and durability testing; - extraction tests for the ReChord System threads and disposable holder; - biocompatibility testing; - MRI compatibility testing; - simulated distribution testing; - package integrity and microbial barrier test after accelerated ageing; - simulated use testing.",
       "mark": [
        64,
        67
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K180411",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "LEGION Knee System",
     "meta": [
      "Smith & Nephew",
      "5 March 2018",
      "K180334",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Mechanical testing To further support a determination of substantial equivalence, a Finite Element Analysis and engineering rationale were conducted on the proposed LEGION Finned Tibial Wedges.",
       "mark": [
        84,
        98
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K180334",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "RTS Lesser MTP Implant System",
     "meta": [
      "In2bones USA",
      "23 February 2018",
      "K173491",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Engineering analysis including Finite Element Analysis was also performed.",
       "mark": [
        31,
        45
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K173491",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Optima Coil System",
     "meta": [
      "Blockade Medical, LLC (D.B.A. Balt Usa)",
      "18 February 2018",
      "K172390",
      "Neurology"
     ],
     "tags": [
      {
       "label": "MRI RF and gradient heating simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "The SEMCAD software package was used to evaluate surface heating patterns for the Opt ima Coil when placed inside the gel of the ASTM phamtom under 1.5T and 3 -T MRI systems.",
       "mark": [
        4,
        10
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K172390",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "The Plateau-LO Spacer System",
     "meta": [
      "Life Spine",
      "2 February 2018",
      "K172105",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance Data: Engineering rationale, finite element analysis and expulsion testing was presented to demonstrate the substantial equivalency of The Plateau-LO Spacer System.",
       "mark": [
        41,
        55
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K172105",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Panthera Dental Milled Bars",
     "meta": [
      "Panthera Dental",
      "26 January 2018",
      "K173466",
      "Dental"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "FEA Finite Element Analysis (FEA) was done on the dental bar prior to starting any testing.",
       "mark": [
        0,
        3
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K173466",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "16ch T/R Hand Wrist Coil",
     "meta": [
      "Quality Electrodynamics",
      "9 January 2018",
      "K173469",
      "Radiology"
     ],
     "tags": [
      {
       "label": "MRI RF and gradient heating simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "A finite-difference time-domain electromagnetic simulation was performed to provide data supporting that the partial body limits for SAR are controlled within the limits described in IEC 60601-2-33.",
       "mark": [
        2,
        31
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K173469",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "EVOS Small Fragment Upper Extremity Plates",
     "meta": [
      "Smith & Nephew",
      "8 January 2018",
      "K173293",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Summary of Pre-Clinical Testing x Finite element analysis (FEA) was conducted on the proposed plate designs to determine the worst case plates based upon structural strength for further mechanical testing.",
       "mark": [
        34,
        48
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K173293",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "PreOp",
     "meta": [
      "Epilog",
      "8 January 2018",
      "K172858",
      "Neurology"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "There is no essential difference in technological characteristics between PreOp and the K001781 device since the Finite Element Model is equivalent to the Finite Difference Model from an individualized forward modeling point of view.",
       "mark": [
        113,
        127
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K172858",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Juliet Ti LL",
     "meta": [
      "Spineart",
      "3 January 2018",
      "K173702",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "A Finite Element Analysis has been submitted to support substantial equivalence.",
       "mark": [
        2,
        16
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K173702",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "AltiVate Anatomic to Reverse Conversion Module",
     "meta": [
      "Encore Medical",
      "26 December 2017",
      "K173073",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Key Differences in Subject Device to Predicate Device(s): Insert length, shell length, gage point, taper and shell fillets Non-Clinical Testing: FEA analysis for stress analysis has determined that the subject device is similar to the evaluated predicate device.",
       "mark": [
        145,
        148
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K173073",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Implantable System for Remodulin",
     "meta": [
      "Medtronic",
      "22 December 2017",
      "P140032",
      "General Hospital"
     ],
     "tags": [
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "The clinical data, mathematical modeling, and bench testing indicate that over the expected longevity of the pump, the accuracy ratio will decrease and plateau at approximately 0.8.",
       "mark": [
        19,
        40
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P140032",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Life Spine Foot and Ankle Plating System",
     "meta": [
      "Life Spine",
      "21 December 2017",
      "K172973",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance Data : Finite Element Analysis, Engineering Rationale and Bench Top Validations were included to demonstrate the substantial equivalency of the Life Spine Foot and Ankle Plating System with respect to: torsional strength, bending strength and axial pullout.",
       "mark": [
        19,
        33
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K172973",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Velofix™ TLIF Cage",
     "meta": [
      "U&I",
      "4 December 2017",
      "K172419",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "The following tests were performed: (1) Static compression test according to ASTM F2077 (2) Dynamic compression test according to ASTM F2077 (3) Static compression-shear test according to ASTM F2077 (4) Dynamic compression-shear test according to ASTM F2077 (5) Subsidence test according to ASTM F2267 Additionally, the finite element analysis was performed to aid in the selection of the worst case implants. 3.",
       "mark": [
        320,
        334
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K172419",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "EluNIR™ Ridaforolimus Eluting Coronary Stent System",
     "meta": [
      "Medinol",
      "28 November 2017",
      "P170008",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "FEA evaluated the structural integrity of the EluNIR stent when subjected to the expected loading conditions during stent manufacturing, delivery, implantation and clinical loading.",
       "mark": [
        0,
        3
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P170008",
     "urlLabel": "FDA database entry"
    }
   ],
   "note": "147 of 9,358 authorizations cite a model. 40 shown: whole sentences first, then table text, newest first in each."
  },
  {
   "id": "2019-2021",
   "label": "2019 to 2021: 1.1% of summaries",
   "count": 98,
   "records": [
    {
     "title": "N1 TiUltra TCC Implant System",
     "meta": [
      "Nobel Biocare Services",
      "21 December 2021",
      "K211109",
      "Dental"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "The impact of the changes on clinical performance was assessed by identifying biomechanical aspects of dental system designs which attribute to long-term performance, assessing each of these aspects thoroughly, and conducting a finite element analysis to assess implant loading distribution.",
       "mark": [
        228,
        242
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K211109",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Meticuly Patient-Specific Titanium Mesh Implant",
     "meta": [
      "Meticuly",
      "21 December 2021",
      "K210099",
      "Neurology"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Table 4: Performance testing summary table Test Test method summary Results Comparative device modeling with Finite Element Analysis (FEA) Computational simulation was performed with a compression load in the normal direction on the center of the mesh implant.",
       "mark": [
        109,
        123
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K210099",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Revolution Apex, Revolution Apex Elite, Revolution Apex Plus, Revolution Apex Select,Revolution CT; Revolution CT ES; Revolution CT with Apex edition; Revolution CT ES with Apex edition",
     "meta": [
      "Ge Medical Systems",
      "17 December 2021",
      "K213715",
      "Radiology"
     ],
     "tags": [
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Non-Clinical Testing The performance evaluation testing included use of a cardiac phantom and mathematical modeling to provide technical substantiation of the 0.23 s/rot performance claims on the Revolution Apex Product Line.",
       "mark": [
        94,
        115
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K213715",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Edwards SAPIEN 3 Transcatheter Pulmonary Valve System with Alterra Adaptive Prestent",
     "meta": [
      "Edwards Lifesciences",
      "16 December 2021",
      "P200015S011",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "To investigate the strain distribution in Finite element analysis Pass the Alterra Adaptive Prestent.",
       "mark": [
        42,
        56
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P200015S011",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Vercise PC, Vercise Gevia and Vercise Genus DBS Systems",
     "meta": [
      "Boston Scientific",
      "20 October 2021",
      "P150031S040",
      "Neurology"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Finite element models were constructed to calculate the electric fields produced by DBS leads in a homogenous tissue medium, and the calculated voltages in space were applied to non-linear axon models in accordance with methods described in the literature for evaluation of DBS electrode design and characterization of field shape [1,2].",
       "mark": [
        0,
        14
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P150031S040",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Portico Transcatheter Aortic Valve Implantation System",
     "meta": [
      "ABBOTT MEDICAL",
      "17 September 2021",
      "P190023",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Visual FEA was used to characterize the Leaflet, Cuff, and structural behavior of the components of In Plane principle stent Finite Element the Portico valve using computer Pass stress Analysis (FEA) analytics when subjected to anticipated in vivo operational conditions Stent Testing Stent Fatigue Freedom from Demonstrate the Portico stent has Pass Resistance fracture fatigue resistance to 600 million cycles.",
       "mark": [
        7,
        10
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P190023",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "RelayPro Thoracic Stent-Graft System",
     "meta": [
      "Bolton Medical",
      "5 August 2021",
      "P200045",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "An evaluation of the potential impact on device performance was conducted that evaluated materials and components, manufacturing processes, anatomical data, biomechanical analysis, computational modeling, bench top performance testing of fractured test samples, and as-applied in-vitro fatigue testing displacements per test article.",
       "mark": [
        181,
        203
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P200045",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Lateral Spine Truss System (LSTS) Interbody Fusion Device",
     "meta": [
      "4Web Medical",
      "5 August 2021",
      "K211388",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance Standards: Validated finite element analysis demonstrated that the product line extension for the Lateral Spine Truss System (LSTS) Interbody Fusion Device does not introduce a new worst-case compared to the previously cleared 4WEB Lumbar Interbody Fusion Devices for mechanical properties of the device.",
       "mark": [
        33,
        47
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K211388",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "POLARSTEM Cemented Femoral Stem",
     "meta": [
      "Smith & Nephew",
      "22 July 2021",
      "K203175",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Non -Clin ical P erformance • Finite Element Analysis per ISO 7206 -4:2010 and ISO 7206 -6:2013.",
       "mark": [
        30,
        44
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K203175",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Pantheon IBFD",
     "meta": [
      "Pantheon Spinal",
      "7 July 2021",
      "K203003",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "An FEA was created to verify that a new worse case was not created with the new size additions.",
       "mark": [
        3,
        6
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K203003",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Neuro Cochlear Implant System",
     "meta": [
      "Oticon Medical",
      "23 June 2021",
      "P200021",
      "Ear, Nose, Throat"
     ],
     "tags": [
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Following clinical and worst-case scenarios, all safety aspects of the stimulation associated with the electrical pulse shape, maximum charge and charge density, leakage or balance between anodic and cathodic phases were compared, evaluated, and validated from computational models and direct implant tests.",
       "mark": [
        261,
        281
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P200021",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "EchoTip Ultra Endoscopic Ultrasound Needle, EchoTip ProCore HD Ultrasound Biopsy Needle, EchoTip Ultra Endobronchial High Definition Ultrasound Needle, EchoTip ProCore Endobronchial High Definition Ultrasound Biopsy Needle",
     "meta": [
      "Cook Ireland",
      "20 May 2021",
      "K210476",
      "Gastroenterology, Urology"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance testing was performed as per Cook Ireland’s design control system; the following tests were conducted: - Tensile testing - Crumple testing - Joint strength testing - Finite element analysis - Simulated use and drop testing VIII.",
       "mark": [
        178,
        192
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K210476",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Persona Revision Knee System Femoral Metaphyseal Cones",
     "meta": [
      "Zimmer",
      "18 May 2021",
      "K210829",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Summary of Performance Data (Nonclinical and/or Clinical) Finite Element Analysis (FEA) was used to predict the peak stress under physiological loading for the subject Persona Revision Knee System femoral metaphyseal cones.",
       "mark": [
        58,
        72
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K210829",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Apex Knee System",
     "meta": [
      "Omnilife Science",
      "2 April 2021",
      "K201611",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Tibial Baseplate: In addition to the testing performed for the Apex Knee System, non-clinical testing specifically for the Pegged Porous Coated Cementless Tibial Baseplates include FEA simulation and peg location templating study.",
       "mark": [
        181,
        184
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K201611",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Harmony™ TPV System",
     "meta": [
      "Medtronic",
      "26 March 2021",
      "P200046",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "To characterize the structural behavior of the For Finite element Harmony TPV frame TPV 22 characterization analysis (FEA) under in vivo operational only. conditions.",
       "mark": [
        51,
        65
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P200046",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Hexanium TLIF",
     "meta": [
      "Spinevision, S.A.S.",
      "10 March 2021",
      "K210359",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance Data The mechanical testing (according to ASTM 2267-04 and ASTM F2077- 14) performed and submitted to support 510(k) clearance of the Hexanium TLIF system (K180437) also supports the subject Hexanium TLIF cages as demonstrated by mechanical and bone fusion worst-case FEA.",
       "mark": [
        280,
        283
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K210359",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "EVOS Large Fragment Plating System",
     "meta": [
      "Smith & Nephew",
      "27 January 2021",
      "K203405",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Summary of Pre-Clinical Testing • Finite element analysis (FEA) was conducted on the proposed plate designs to determine the worst case plates for further mechanical testing.",
       "mark": [
        34,
        48
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K203405",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "CarboClear Lumbar Cage System",
     "meta": [
      "Carbofix Orthpedics",
      "15 January 2021",
      "K203683",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance characteristics included finite element analysis with supportive testing according to ASTM F2077 and expulsion test.",
       "mark": [
        37,
        51
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K203683",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "SIRION Lateral Lumbar Interbody Fusion",
     "meta": [
      "Astura Medical",
      "6 January 2021",
      "K202495",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Non-clinical Test The following analyses were conducted: Summary • FEA o Static and Dynamic Compression (ASTM F2077) o Static and Dynamic Shear (ASTM F2077) o Subsidence (ASTM F2267) • Mechanical o Expulsion The results of these evaluations indicate that the Sirion implants are equivalent to predicate devices.",
       "mark": [
        67,
        70
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K202495",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Vercise Deep Brain Stimulation (DBS) System",
     "meta": [
      "Boston Scientific",
      "29 December 2020",
      "P150031S028",
      "Neurology"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Finite element models were constructed to calculate the electric fields produced by DBS leads in a homogenous tissue medium, and the calculated voltages in space were applied to non- linear axon models in accordance with methods described in the literature for evaluation of DBS electrode design and characterization of field shape [1,2].",
       "mark": [
        0,
        14
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P150031S028",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "OPRA Implant System",
     "meta": [
      "Integrum",
      "18 December 2020",
      "P190009",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "OPRATM Fatigue Testing wear and corrosion fatigue life at 10M 1.2 showed less wear than resistance for cycles of ±15 Nm OPRATM 1.0 after 10M intended use life in than OPRATM 1.0 cycles. saline 3 samples Laboratory and Simulated Use Studies to Support Changes since Original HDE Approval for the OPRATM Implant System FEM Modeling for Comparative Finite element Lower stress Passed.",
       "mark": [
        346,
        360
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P190009",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "HydroMARK Breast Biopsy Site Markers",
     "meta": [
      "Devicor Medical Products",
      "16 December 2020",
      "K203097",
      "General, Plastic Surgery"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "The FEA Simulation study is based on design requirements of the product and are unique to this product type, there is no recognized testing standard.",
       "mark": [
        4,
        7
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K203097",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "EVOS Large Fragment Plating System",
     "meta": [
      "Smith & Nephew",
      "16 November 2020",
      "K201918",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Summary of Pre-Clinical Testing • Finite element analysis (FEA) was conducted on the proposed plate designs to determine the worst case plates for further mechanical testing.",
       "mark": [
        34,
        48
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K201918",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Acumen Assisted Fluid Management (AFM) Software Feature",
     "meta": [
      "Edwards Lifesciences",
      "13 November 2020",
      "DEN190029",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Monte Carlo simulation"
      }
     ],
     "quotes": [
      {
       "label": "FDA's decision summary",
       "text": "For each iteration of the Monte Carlo simulation, the initial volume in the IV bag was chosen from a uniform distribution between 500 mL and b(x) and the simulation decremented this volume after each simulated bolus to simulate the delivery of fluid boluses from a b(x) IV bag.",
       "mark": [
        26,
        37
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/denovo.cfm?id=DEN190029",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Unid IB3D Alif",
     "meta": [
      "Medicrea International",
      "30 October 2020",
      "K200316",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Mechanical testing Finite Element Analysis (FEA) was conducted and it was determined that no new worst-case implant was created.",
       "mark": [
        19,
        33
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K200316",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Abre Venous Self-expanding Stent System",
     "meta": [
      "Medtronic Vascular",
      "21 October 2020",
      "P200026",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Finite Element Characterization testing performed to determine maximum stresses and strains within Analysis the device to support fatigue analysis.",
       "mark": [
        0,
        14
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P200026",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "TruForm Sievers Annuloplasty Ring",
     "meta": [
      "Genesee Biomedical",
      "6 October 2020",
      "K202253",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Testing was performed to support substantial equivalence, including: Biocompatibility Sterilization Computational Finite Element Analysis (FEA) Tensile strength of ring materials Suture pull-out testing MRI Compatibility Packaging Shelf-life Conclusion Genesee BioM edical, Inc. considers the TruForm™ Sievers Annuloplasty Ring to be equivalent to the predicate device.",
       "mark": [
        114,
        128
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K202253",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Talos®-C Cervical Intervertebral Body Fusion System, Talos®-C (HA) Cervical Intervertebral Body Fusion System",
     "meta": [
      "Meditech Spine",
      "11 September 2020",
      "K201506",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Finite Element Analysis and engineering analysis were performed on the subject devices to confirm that a new worst‐case is not presented in the proposed devices.",
       "mark": [
        0,
        14
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K201506",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Diamond Digital Sleep Orthotic (DDSO)",
     "meta": [
      "Diamond Orthotic Laboratory",
      "3 September 2020",
      "K202499",
      "Dental"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Validation activities include, but are not limited to: usability studies, finite element analysis validated through mechanical testing, performance testing and several comparative analysis in parallel with the predicate device.",
       "mark": [
        74,
        88
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K202499",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Spartan Hip Stem",
     "meta": [
      "Signature Orthopaedics",
      "5 August 2020",
      "K192883",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Non-clinical testing carried out included: • Range of Motion testing as per ISO 21535 • Stem and Neck Fatigue FEA • Stem and Neck Fatigue Testing • CoCr Head Disassembly Test • Ceramic Head Static Burst Test, Fatigue and Post-Fatigue Burst Test • Ceramic Head Pull-Off and Torsion Test The subject device has been validated as non-pyrogenic.",
       "mark": [
        110,
        113
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K192883",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "EVOS Straight Proximal Humerus 7-15 hole Plates (137-225mm)",
     "meta": [
      "Smith & Nephew",
      "22 July 2020",
      "K201527",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Based on the similarities to the predicate devices and a review of the finite element analysis testing , the subject device is substantially equivalent to the commercially available predicate device s listed above.",
       "mark": [
        71,
        85
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K201527",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "WATCHMAN FLX Left Atrial Appendage Closure Device with Delivery System and WATCHMAN Left Atrial Appendage Closure Device",
     "meta": [
      "Boston Scientific",
      "21 July 2020",
      "P130013S035",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Finite Element This test assessed the stress/strain levels in Pass Analysis (for key areas of the implant characterization) Implant Durability Implant Durability/Fatigue (Test articles Pass were cycled up to the equivalent of 10 years of use or 400,000,000 cycles) These tests quantitatively evaluate resistance to in- vivo fatigue-related damage.",
       "mark": [
        0,
        14
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P130013S035",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "AltiVate Anatomic Canal-Sparing (CS) Shoulder",
     "meta": [
      "Encore Medical",
      "3 June 2020",
      "K193226",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "This testing includes: • Taper Disassociation Testing • Humeral Construct Fatigue Testing - Torque-Out Testing • Axial Pull-Out Testing • Humeral Micromotion - Cadaveric Experimental Analysis • Humeral Micromotion - Finite Element Analysis (FEA) • Humeral Fit Analysis All testing has determined that the device is substantially equivalent to the predicate devices.",
       "mark": [
        216,
        230
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K193226",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "LV Liberty Vision Model 1 90Yttrium Brachytherapy Source",
     "meta": [
      "Lv Liberty Vision",
      "29 May 2020",
      "K193602",
      "Radiology"
     ],
     "tags": [
      {
       "label": "Monte Carlo simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Dosimetry The dose distribution around the LV Liberty Vision Model 1 90Yttrium Brachytherapy Source was calculated by Monte Carlo simulation.",
       "mark": [
        118,
        129
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K193602",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "NuVasive® ACP System",
     "meta": [
      "Nu Vasive, Incorporated",
      "20 May 2020",
      "K201078",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "The following testing was performed: x Static torsion\u000f\u0003VWDWLF\u0003FRPSUHVVLRQ\u0003EHQGLQJ\u000f\u0003DQG\u0003G\\QDPLF\u0003FRPSUHVVLRQ\u0003EHQGLQJ testing per ASTM F1717 In addition to the testing above, Finite Elemental Analysis and engineering rationales were used to establish that the minor design updates introduced did not create a new worst case.",
       "mark": [
        172,
        186
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K201078",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "TREO® Abdominal Stent-Graft System",
     "meta": [
      "Bolton Medical",
      "4 May 2020",
      "P190015",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "A root cause analysis and evaluation of the potential impact on device performance was conducted that evaluated materials and components, manufacturing processes, clinical procedure, patient anatomical data, bench top performance testing of fractured test samples, biomechanical analysis, computational modeling, and experimental fatigue testing.",
       "mark": [
        289,
        311
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P190015",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Sonata® Sonography-Guided Transcervical Fibroid Ablation System 2.1",
     "meta": [
      "Gynesonics",
      "4 May 2020",
      "K193516",
      "Obstetrics/Gynecology"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Effect of Parameter Variations on Ablation Dimensions - Modeling demonstrates ex vivo non- Finite Element Method (FEM) perfused tissue model is conservative Variation in Computational Modeling to determine model to establish AZ and TSB.",
       "mark": [
        91,
        105
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K193516",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Evry",
     "meta": [
      "Synaptive Medical",
      "29 April 2020",
      "K200327",
      "Radiology"
     ],
     "tags": [
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Non-clinical testing included computational modeling of the integrated RF coil and gradient system demonstrating equivalence to the predicate device with respect to cardiac stimulation and RF energy deposition.",
       "mark": [
        30,
        52
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K200327",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Panthera Dental Milled Bars",
     "meta": [
      "Panthera Dental",
      "8 April 2020",
      "K192108",
      "Dental"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Therefore , we believe that the new device is as safe as the legally marketed device. 5.5.3 FEA Finite Element Analysis (FEA) was done on the bar cylinders prior to starting any testing. 5.5.4 Fatigue Testing Fatigue testing was perfo rmed on the Panthera Dental bar .",
       "mark": [
        92,
        95
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K192108",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Neural Navigator",
     "meta": [
      "Soterix Medical",
      "6 April 2020",
      "K191422",
      "Neurology"
     ],
     "tags": [
      {
       "label": "Monte Carlo simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "It is tested via: maximum) 1) Monte Carlo simulations of the mapping algorithm and the ensuing navigation are run 10,000 times using realistic position measurement noise conditions. 2) Clinical study using the CloudTMS volunteers).",
       "mark": [
        30,
        41
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K191422",
     "urlLabel": "FDA database entry"
    }
   ],
   "note": "98 of 9,017 authorizations cite a model. 40 shown: whole sentences first, then table text, newest first in each."
  },
  {
   "id": "2022-2023",
   "label": "2022 to 2023: 0.9% of summaries",
   "count": 57,
   "records": [
    {
     "title": "Occlutech® ASD Occluder and Occlutech® Pistol Pusher",
     "meta": [
      "Occlutech Holding",
      "29 December 2023",
      "P200032",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Occlutech ASD Occluder Table 2: Occlutech ASD Occluder Functional, Shelf Life, Biocompatibility, and Sterility Testing Test Suite Test Description Summary of Results Bench Tests Device Durability 400 million cycles of device durability testing Passed Finite Element Safety factor >1 Passed Analysis Fatigue strength of To characterize the fatigue strength of the Characterization only.",
       "mark": [
        251,
        265
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P200032",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "KeyLift™ Expandable Interlaminar Stabilization System",
     "meta": [
      "Amplio Spine",
      "19 October 2023",
      "K232484",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "W orst-case implants were identified using FEA and/or engineering analyses.",
       "mark": [
        43,
        46
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K232484",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "TriLEAP Plating System",
     "meta": [
      "Crossroads Extremity Systems",
      "29 September 2023",
      "K230591",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance Data • Plate Cross-Section Analysis (Bending Strength) • Screw Torsional Strength Analysis according to ASTM F543 via a Finite Element Analysis model • Screw Driving Torque according to ASTM F543 • Screw Pull - Out via the Chapman Equation Clinical Clinical testing was not necessary for the determination of substantial Performance Data equivalence.",
       "mark": [
        132,
        146
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K230591",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "LimFlow System",
     "meta": [
      "LimFlow",
      "11 September 2023",
      "P220025",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "The calculation of Element Analysis Safety Factor (SF) and (FEA) determination of worst case conditions for accelerated durability testing was also performed The implant must exhibit acceptable To evaluate the durability simulated 10-year durability.",
       "mark": [
        60,
        63
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P220025",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Minitouch 3.8 Era System (Minitouch System)",
     "meta": [
      "Microcube",
      "28 July 2023",
      "P230002",
      "Obstetrics/Gynecology"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Thermal Modeling The applicant reported the results of finite element simulation of the thermal performance of the Minitouch System.",
       "mark": [
        0,
        16
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P230002",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "iTFlow",
     "meta": [
      "Cardio Flow Design",
      "12 May 2023",
      "K222854",
      "Radiology"
     ],
     "tags": [
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Numerical phantom test result The software algorithms are validated by comparing with computer simulation.",
       "mark": [
        86,
        105
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K222854",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Breast Implant Removal Device (BIRD)",
     "meta": [
      "Gaylord Solutions",
      "20 April 2023",
      "DEN220082",
      "General, Plastic Surgery"
     ],
     "tags": [
      {
       "label": "In silico or virtual patients"
      }
     ],
     "quotes": [
      {
       "label": "FDA's decision summary",
       "text": "The minimum and maximum recommended pressure settings were used. - Simulated Use/Usability Testing used the same in silico breast model as functional testing.",
       "mark": [
        113,
        122
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/denovo.cfm?id=DEN220082",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "3M V.A.C. Veraflo Cleanse Choice Dressing Kit, 3M Veraflo Cleanse Choice Complete Dressing Kit",
     "meta": [
      "3M Healthcare Business Group",
      "23 March 2023",
      "K221585",
      "General, Plastic Surgery"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "The testing consisted of the following: o A comparison of materials of construction o A comparison of material physical properties, such as pore size, density, tensile, tear, elongation, compression, hole size, hole pattern, and geometry o Finite element modeling of expected strain profiles o Foam changes under negative pressure as measured by surface area and hole size contraction.",
       "mark": [
        240,
        254
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K221585",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "iCast Covered Stent System",
     "meta": [
      "Atrium Medical",
      "22 March 2023",
      "P120003",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "To determine maximum Strains were calculated using the Stress/Strain PASS stresses and strains within worst-case configurations in a Analysis the device. validated FEA model. )DWLJXH\u00036DIHW\\\u0003)DFWRU\u0003\u0003\u0014\u0011\u0013\u0011 All data points on Goodman To evaluate the durability Fatigue Diagram for stent at maximum PASS and integrity of the stent via Analysis diameter must be below the finite element analysis.",
       "mark": [
        367,
        381
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P120003",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "SPIRA®-C Integrated Fixation System",
     "meta": [
      "Camber Spine Technologies",
      "22 March 2023",
      "K223837",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Testing, through the use of Finite Element Analysis (FEA) simulations, shows that the anchors are equivalent to the screw -based construct and has less internal stress than the non-integrated SPIRA® -C Open Matrix cage.",
       "mark": [
        28,
        42
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K223837",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "MagVenture TMS Atlas Neuro Navigation System",
     "meta": [
      "Tonica Elektronik",
      "17 March 2023",
      "K221544",
      "Neurology"
     ],
     "tags": [
      {
       "label": "Monte Carlo simulation"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Monte Carlo simulations of the mapping algorithm and the ensuing navigation precision ran 10.000 times, using realistic position measurements noise conditions (as obtained from literature).",
       "mark": [
        0,
        11
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K221544",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Magstim Horizon 3.0 TMS Therapy System, Horizon 3.0 System, Horizon 3.0, H3.0, Horizon 3.0 with StimGuide+",
     "meta": [
      "Magstim",
      "16 March 2023",
      "K223154",
      "Neurology"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "The Horizon 3.0 Ez Cool Coil and the NeuroStar Advanced Therapy Coil electric field profiles have also been modeled in COMSOL and according to the finite element modelling (FEM) and statistical analysis, the induced E- field profiles by the Horizon 3.0 Ez Cool Coil and Neuro Star Advanced Therapy Coil are equivalent and have an average difference of ±5% at distances of 1 to 4 cm from the coil surface.",
       "mark": [
        119,
        125
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K223154",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Injection PIN (KIP(02031-02061) (03031-03061))",
     "meta": [
      "Slk Ortho",
      "3 March 2023",
      "K221697",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "In silico or virtual patients"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "In silico simulation in Labs of Politecnico di Milano was validated to perform 4 different scenarios in comparison with predicate device V-Strut.",
       "mark": [
        0,
        9
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K221697",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "RefleXion Medical Radiotherapy System (RMRS)",
     "meta": [
      "Reflexion Medical",
      "1 February 2023",
      "DEN220014",
      "Radiology"
     ],
     "tags": [
      {
       "label": "In silico or virtual patients"
      }
     ],
     "quotes": [
      {
       "label": "FDA's decision summary",
       "text": "LIMITATIONS - The sale, distribution, and use of the RMRS are restricted to prescription use in accordance with 21 CFR 801.109. - The clinical study did not include treatment of patients with the RMRS device, and instead assessed effectiveness through the use of data simulated in silico. - The quality of the positron emission tomography (PET) images obtained using the RMRS device are not of diagnostic quality.",
       "mark": [
        278,
        287
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/denovo.cfm?id=DEN220014",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Magstim Horizon 3.0 TMS Therapy System, Horizon 3.0 System, Horizon 3.0, H3.0, Horizon 3.0 with StimGuide+",
     "meta": [
      "Magstim",
      "13 January 2023",
      "K222171",
      "Neurology"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "In addition to OCD protocol performance testing, the subject device (Horizon 3.0) and the primary predicate device (NeuroStar) applicating coils were modeled in COMSOL and according to Finite Element Method (FEM) and statistical analysis, the induced E-field profiles are equivalent and have an average difference of ±5% at distances of 1 to 4 cm from the coil surface.",
       "mark": [
        161,
        167
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K222171",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Cervical Spine Truss System (CSTS) Interbody Fusion Device",
     "meta": [
      "4Web Medical",
      "5 January 2023",
      "K221266",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance Standards: Validated finite element analysis demonstrated that the product line extension for the Cervical Spine Truss System (CSTS) does not introduce a new worst-case compared to the previously cleared 4WEB Cervical Spine Truss System-Stand Alone (CSTS-SA) devices for mechanical performance of the device.",
       "mark": [
        33,
        47
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K221266",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "aprevo® Digital Workflow",
     "meta": [
      "Carlsmed",
      "30 December 2022",
      "K222195",
      "Radiology"
     ],
     "tags": [
      {
       "label": "In silico or virtual patients"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "In addition, the measuring function of the Subject Device was evaluated in bench testing using in-silico phantoms with known values.",
       "mark": [
        95,
        104
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K222195",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "EVOS Wrist Spanning Plate",
     "meta": [
      "Smith & Nephew",
      "18 October 2022",
      "K213123",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Summary of Pre-Clinical Testing • Bending performance was evaluated using a four-point bend fatigue test identified through FEA as the worst-case scenario in accordance with the guidelines of ASTM F382.",
       "mark": [
        124,
        127
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K213123",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Resolute Onyx Zotarolimus-Eluting Coronary Stent System, Onyx Frontier Zotarolimus-Eluting Coronary Stent System",
     "meta": [
      "Medtronic Vascular",
      "15 September 2022",
      "P160043S058",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "New preclinical testing performed to support the use of the stent in non-left main bifurcation lesions treated using a provisional stenting strategy included FEA analysis, coating durability, and acute particulate evaluation.",
       "mark": [
        158,
        161
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P160043S058",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "PASCAL Precision Transcatheter Valve Repair System",
     "meta": [
      "Edwards Lifesciences",
      "14 September 2022",
      "P220003",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Table 3: Summary of PASCAL Precision System Bench Testing Test Purpose Results PASCAL and PASCAL Ace Implants No fracture of implant To determine mechanical structural components stress/strain during device predicted within a minimum Finite element analysis loading, deployment and cyclic of 600 million cycles under loading.",
       "mark": [
        234,
        248
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P220003",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Hammertoe Truss System (HTS)",
     "meta": [
      "4Web Medical",
      "6 September 2022",
      "K221283",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Page2 of 2 Performance Standards: Validated finite element analysis demonstrated that the additional implant sizes for the Hammertoe Truss System (HTS) do not create a new worst-case implant compared to the previously cleared 4WEB Hammertoe Truss System devices for mechanical properties of the device.",
       "mark": [
        44,
        58
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K221283",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix)",
     "meta": [
      "I.T.S.",
      "31 August 2022",
      "K210935",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance testing as a consensus standard was performed under ASTM F543 for all bone screws and under ASTM F382 for bone plates and in addition a FEA analysis simulation following ASTM F382.",
       "mark": [
        148,
        151
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K210935",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "SmartBolus Calculator",
     "meta": [
      "Insulet",
      "19 August 2022",
      "K222239",
      "Clinical Chemistry"
     ],
     "tags": [
      {
       "label": "In silico or virtual patients"
      }
     ],
     "quotes": [
      {
       "label": "FDA's decision summary",
       "text": "In Silico Testing: Simulations of possible input and output combinations were run to supplement information collected in the Clinical Study.",
       "mark": [
        0,
        9
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K222239",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "iStent infinite Trabecular Micro-Bypass System, Model iS3",
     "meta": [
      "Glaukos",
      "2 August 2022",
      "K220032",
      "Ophthalmic"
     ],
     "tags": [
      {
       "label": "Computational fluid dynamics"
      },
      {
       "label": "Finite element analysis"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Pressure/Flow Characteristics Numerical modeling, including computational fluid dynamics, was used to evaluate the flow through the stents over physiologically relevant boundary conditions.",
       "mark": [
        30,
        48
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K220032",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "VyPlate Anterior Cervical Plate System",
     "meta": [
      "Vy Spine",
      "26 July 2022",
      "K221572",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance Data Design information and finite element analysis were utilized to establish that the subject VyPlate™ Anterior Cervical Plates have substantially equivalent mechanical strength to the predicate Reliance Anterior Cervical Plate System (K122216, K140742).",
       "mark": [
        40,
        54
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K221572",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "KLS Martin IPS Preprosthetic",
     "meta": [
      "KLS-Martin",
      "8 July 2022",
      "K210228",
      "Dental"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Fourteen (14) Selective Laser Melting (SLM) test articles, identical to the final finished device with regard to processing, were used to examine the influence of various bed positions and any effect of component dimensional accuracy using FEM analysis.",
       "mark": [
        240,
        243
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K210228",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "CerroZone Mobile",
     "meta": [
      "Cerrozone",
      "1 July 2022",
      "K220298",
      "General Hospital"
     ],
     "tags": [
      {
       "label": "Computational fluid dynamics"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Determination of Turbulent Air Flow and/or Site Contamination: Risk of turbulent air flow and/or potential site contamination in areas that have a sterile field or controlled air flow has been assessed via computational fluid dynamics per ASHRAE/ASHE Standard 170-2017 Ventilation of Health Care Facilities.",
       "mark": [
        206,
        233
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K220298",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "GORE TAG Thoracic Branch Endoprosthesis (TBE Device)",
     "meta": [
      "W. L. Gore & Associates",
      "13 May 2022",
      "P210032",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Branch Endoprosthesis components must Fatigue Finite Element withstand simulated physiologic pulsatile Analysis was used to loading for ten years without wire fractures, determine the strains failure of the graft material, or failure of the present in the wire stent graft material/Nitinol endoprosthesis structure under specific composite that would compromise device loading conditions for function. each size.",
       "mark": [
        46,
        60
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P210032",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "BioFreedom Drug Coated Coronary Stent System",
     "meta": [
      "Biosensors International USA",
      "14 April 2022",
      "P190020",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Minimum compression resistance of 0.1 N/mm of stent length at 0.5 mm of stent compression Stress/Strain To determine the stent durability N/A N/A and Fatigue when exposed to worst case Analysis physiological loads and configuration by means of a Finite Element Analysis.",
       "mark": [
        246,
        260
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P190020",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Ri.hip Modeler",
     "meta": [
      "Blue Belt Technologies",
      "11 March 2022",
      "K212040",
      "Radiology"
     ],
     "tags": [
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "The simulated model is derived from validated computational models that effectively display the range of implant motion in the model.",
       "mark": [
        46,
        66
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K212040",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Genesys Spine 3DP Lumbar Interbody System",
     "meta": [
      "Genesys Spine",
      "9 March 2022",
      "K220096",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Finite Element Analysis (FEA) surface area analyses, graft space evaluations, and risk assessments were performed and lead to the conclusion that the 3DP Lumbar Interbody Line Extension additions are substantially equivalent to the Genesys Spine 3DP Lumbar Interbody System (Genesys Spine – K182987).",
       "mark": [
        0,
        14
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K220096",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Genesys Spine 3DP AIS-C II Cervical Interbody System",
     "meta": [
      "Genesys Spine",
      "13 December 2023",
      "K233594",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "…Interbody: In order to serve a larger range of the population and user demographic an explicit 10 degree lordotic interbody was created. o Finite Element Analysis was conducted, and the results proved that the 10 degree lorodtic interbody did not introduce any new worst-case in regards to mechanical stresses and standardized testing and was still superior to predicate. o Design Validation testing was performed, and…",
       "mark": [
        140,
        154
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K233594",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Arthrex Univers Apex OptiFit Humeral Stem",
     "meta": [
      "Arthrex",
      "26 September 2023",
      "K230513",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "MRI RF and gradient heating simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Additional computational electromagnetic simulation testing per FDA guidance, “Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment,” to assess RF heating in bone tissue demonstrated that the Arthrex Shoulder Systems are “MR Conditional.” Technological Comparison The proposed device has the same technological characteristics (material, sterilization method and biocompatibility…",
       "mark": [
        25,
        51
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K230513",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Cervical Spine Truss System (CSTS)",
     "meta": [
      "4Web Medical",
      "25 September 2023",
      "K231739",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "…– Dynamic Compression Shear (With and Without the Integrated Plate) • ASTM F2077 – Dynamic Torsion (With and Without the Integrated Plate) - Screw Pushout Testing • Expulsion Testing per accepted industry standard Validated finite element analysis demonstrated that the worst case CSTS device for each endplate geometry is not a new worst case compared to the previously tested CSTS devices within the 4WEB portfolio.",
       "mark": [
        225,
        239
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K231739",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "BlueStar CGM insulin dose calculator",
     "meta": [
      "Welldoc",
      "11 August 2023",
      "K222888",
      "Clinical Chemistry"
     ],
     "tags": [
      {
       "label": "In silico or virtual patients"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "FDA's decision summary",
       "text": "…1.9 < 70 0.98 [1.07] 0.79 [0.81] -0.19 < 54 0.18 [0.28] 0.12 [0.19] -0.06 > 180 32.15 [12.66] 30.41 [17.49] -1.74 ≥ 250 5.85 [5.11] 6.27 [7.47] 0.42 ≥ 300 1.30 [1.87] 1.66 [3.05] 0.36 In-Silico Testing: A mathematical simulation of the possible input combinations as well as a simulated clinical study using virtual patients were used to assess potential edge cases (including physiologically improbable edge cases).",
       "mark": [
        185,
        194
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K222888",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Miria Skin Treatment System",
     "meta": [
      "AVAVA",
      "22 June 2023",
      "K223871",
      "General, Plastic Surgery"
     ],
     "tags": [
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "…Part 1: General requirements for basic safety  EMC Testing was conducted per IEC 60601-1-2: 2014-02 Medical electrical equipment - Part 1-2 General requirements for basic safety and essential performance  Histology study and computational simulations were conducted to evaluate lesion geometry  Sample clinical data were provided to confirm the shape of the CTZs (conical thermal zones) and demonstrate safety.",
       "mark": [
        228,
        253
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K223871",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Xerxes Hip Stem",
     "meta": [
      "Xerxes Arthopedix",
      "15 June 2023",
      "K231109",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Non-clinical testing carried out included: • Range of Motion testing as per ISO 21535 • Stem and Neck Fatigue FEA as per ASTM F2996 • Stem and Neck Fatigue Testing as per ISO 7206-4 and ISO 7206-6 • Pyrogenicity and Endotoxin Testing as per AAMI ST72 Conclusion: XerxesTM Stem is substantially equivalent to Medacta International’s MiniMAX Stem (K170845 and K192352) since it has the same intended use, indications for…",
       "mark": [
        110,
        113
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K231109",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "DETOUR System",
     "meta": [
      "Endologix",
      "7 June 2023",
      "P220021",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "To evaluate the finite element FEA is used to determine the Worst case sizes Finite Element analysis (FEA) results to the worst-case stent graft sizing for for accelerated Analysis (FEA) self-expanding Torus Stent Graft durability fatigue testing. durability testing under simulated in vivo pulsatile Therefore, there is no pre- were identified and multi-mode physiological defined acceptance criteria for loading…",
       "mark": [
        16,
        30
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P220021",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "RelayPro Thoracic Stent-Graft System",
     "meta": [
      "Bolton Medical",
      "7 March 2023",
      "P200045S002",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Pass Durability – analysis (FEA) was The worst-case component Computational used to compute the size was identified and Analyses maximum strains in used to select the worst- all of the RelayPro case prosthesis oversizing design’s sizes when for in-vitro fatigue testing. subjected to catheter loading and Worst-case identification maximum alternating also considered strains in an in-vivo differences in vascular…",
       "mark": [
        28,
        31
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P200045S002",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Elucis",
     "meta": [
      "Realize Medical",
      "17 January 2023",
      "K220649",
      "Radiology"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "…The subject device does not support automated workflows with scripting, nor can it be interfaced with third-party applications for finite element analysis Despite these technological differences, the additional modeling and annotation capabilities available in the subject device are a matter of convenience for users, and do not affect the format of the final output produced by the system, nor the safety and…",
       "mark": [
        132,
        146
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K220649",
     "urlLabel": "FDA database entry"
    }
   ],
   "note": "57 of 6,606 authorizations cite a model. 40 shown: whole sentences first, then table text, newest first in each."
  },
  {
   "id": "2024-2026",
   "label": "2024 to 2026, after FDA's credibility guidance: 1.2% of summaries",
   "count": 110,
   "records": [
    {
     "title": "iOS CamAPS FX",
     "meta": [
      "CamDiab",
      "16 September 2026",
      "K261896",
      "Clinical Chemistry"
     ],
     "tags": [
      {
       "label": "In silico or virtual patients"
      }
     ],
     "quotes": [
      {
       "label": "FDA's decision summary",
       "text": "Once a compatible candidate iCGM is identified, the sponsor will conduct virtual clinical trial testing with virtual patients, with pre-defined acceptance criteria based on virtual patient equivalency testing and glycemic outcomes from the iAGC clinical study.",
       "mark": [
        73,
        95
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K261896",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "OsteoCoil™ GT Active Compression System",
     "meta": [
      "Flower Orthopedics",
      "24 August 2026",
      "K260106",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "The OsteoCoilTM GT Active Compression System and its Predicate Device are significantly similar in design, materials, and manufacturing , and Finite Element Analysis was performed to further confirm expected performance of the device.",
       "mark": [
        142,
        156
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K260106",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "METICULY Patient-specific CMF solution",
     "meta": [
      "Meticuly",
      "25 July 2026",
      "K253206",
      "Dental"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "These characteristic differences were evaluated through comp rehensive non -clinical performance testing, including material characterization, mechanical bench testing, finite element analysis (FEA), and modified compression testing.",
       "mark": [
        169,
        183
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K253206",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "E3D™-C Interbody System",
     "meta": [
      "Evolution Spine",
      "22 July 2026",
      "K262094",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "This conclusion was supported through finite element analysis (FEA), cross-sectional evaluations, and confirmatory mechanical testing of the integrated plate construct, including dynamic shear and dynamic torsion testing performed in accordance with ASTM F2077, as well as wear debris analysis per ASTM F1877.",
       "mark": [
        38,
        52
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K262094",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "MISAGO(TM) RX Self-expanding Peripheral Stent",
     "meta": [
      "Terumo Medical",
      "16 July 2026",
      "P140002S032",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Fatigue Analysis pulsatile loading. the stent using finite element analysis.",
       "mark": [
        52,
        66
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P140002S032",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Perfusion Measurement Platform v1",
     "meta": [
      "Quantified Imaging",
      "22 June 2026",
      "K253421",
      "Radiology"
     ],
     "tags": [
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Because direct testing of the predicate device was not available, non-clinical performance testing was designed to assess the technological differences using representative legally marketed reference device datasets and computational simulation methods.",
       "mark": [
        220,
        244
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K253421",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "MEMO 4D Curve™ Semirigid mitral annuloplasty ring",
     "meta": [
      "Corcym",
      "18 June 2026",
      "K261707",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "The core stiffness is calculated through a Finite Element Method and varies between the core posterior portion (rigid) and the two core ends (flexible), slightly increasing with the core size.",
       "mark": [
        43,
        57
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K261707",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "KT Airway Clearance Device II (KT-3); KT Airway Clearance Device II (KT-5)",
     "meta": [
      "Hunan Ruijiong Biotech",
      "16 June 2026",
      "K253001",
      "Anesthesiology"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Results confirm that the KT Airway Clearance Device II and the predicate have comparable waveforms across the various modes for mechanical insufflation- exsufflation therapy.  Device lifetime evaluated using engineering analysis (Finite Element Analysis).",
       "mark": [
        231,
        245
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K253001",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "MySleepDash",
     "meta": [
      "Somnetics International",
      "17 April 2026",
      "K252338",
      "Anesthesiology"
     ],
     "tags": [
      {
       "label": "Monte Carlo simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Monte Carlo are In addition, therapy data substantially equivalent may be accessed through a with respect to supported web-based interface using a host platforms. supported internet browser, depending on device compatibility and regional availability.",
       "mark": [
        0,
        11
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K252338",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Acorys Mapping System",
     "meta": [
      "Corify Care S.L",
      "13 April 2026",
      "K253861",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "No adverse impact on mapping accuracy was observed due to the differences; indeed, the ACORYS configuration yielded equivalent or improved localization accuracy in mathematical simulation.",
       "mark": [
        164,
        187
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K253861",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "aura6000™ system",
     "meta": [
      "LivaNova USA",
      "18 March 2026",
      "P250013",
      "Anesthesiology"
     ],
     "tags": [
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "This evaluation included anatomical review of structures adjacent to the proximal hypoglossal nerve, computational modeling of electric field distribution including worst-case proximity modeling this adjacency, and assessment of stimulation-related adverse events observed during clinical use.",
       "mark": [
        101,
        123
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P250013",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Rebel Platinum Chromium Coronary Stent System Monorail and Over the Wire",
     "meta": [
      "Boston Scientific",
      "10 March 2026",
      "P130030",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "The FEA evaluated the structural integrity of the over a 10 year stent when subjected to the expected loading conditions evaluation. generated in coronary arteries.",
       "mark": [
        4,
        7
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P130030",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "DePuy Synthes VOLT™ Medial Distal Femur 3.5 Plating System",
     "meta": [
      "Synthes",
      "10 March 2026",
      "K254093",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Non-Clinical To demonstrate the safety and efficacy of the subject devices and Performance support the substantial equivalence to their predicates, the Testing following testing was performed: • Computational Finite Element Analysis simulating a mechanical static bending construct test, demonstrated subject plates were non-inferior regarding their yield load.",
       "mark": [
        209,
        223
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K254093",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "DePuy Synthes VOLT™ Proximal Tibia 3.5 Plating System",
     "meta": [
      "Synthes",
      "9 March 2026",
      "K260069",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Non - Clinical To demonstrate the safety and efficacy of the subject devices and support the Performance Testing substantial equivalence to their predicates, the following testing was performed: • Computational Finite Element Analysis simulating a mechanical static bending construct test, demonstrated subject plates were non - inferior regarding their bending moment and stiffness.",
       "mark": [
        211,
        225
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K260069",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "VOLT™ Ankle Trauma 2.7/3.5 Plating System; VOLT™ Calcaneus 2.7 Plating System",
     "meta": [
      "Synthes",
      "2 March 2026",
      "K254054",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Non-Clinical To demonstrate the safety and efficacy of the subject devices and support the Performance substantial equivalence to their predicates, the following testing was performed: Testing • Computational Finite Element Analysis simulating a mechanical static bending construct test, demonstrated subject plates were non- inferior regarding their bending strength.",
       "mark": [
        209,
        223
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K254054",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI Panvivo S); uMI Panvivo (uMI Panvivo EX); uMI Panvivo (uMI Panvivo ES)",
     "meta": [
      "Shanghai United Imaging Healthcare",
      "13 February 2026",
      "K253564",
      "Radiology"
     ],
     "tags": [
      {
       "label": "Monte Carlo simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "For the Monte Carlo -simulated (MCS) case, we simulated one 500 -second brain scan with the subject completely stationary, and another brain scan with identical simulation settings but with rot ational motion around the neck occurring during the scan.",
       "mark": [
        8,
        19
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K253564",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Libertas Taper Short (TS) Uncemented Femoral Stem (Libertas TS Stem)",
     "meta": [
      "Maxx Orthopedics",
      "15 January 2026",
      "K253171",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "This included finite element analysis to identify worst case sizes for both Distal Fatigue (Stem Body Endurance) and Proximal Fatigue (Neck Endurance).",
       "mark": [
        14,
        28
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K253171",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "SmartGuard technology; Predictive Low Glucose technology",
     "meta": [
      "Medtronic Minimed",
      "14 January 2026",
      "K253585",
      "Clinical Chemistry"
     ],
     "tags": [
      {
       "label": "In silico or virtual patients"
      }
     ],
     "quotes": [
      {
       "label": "FDA's decision summary",
       "text": "For iCGMs, re- qualification requires demonstrated continued compatibility to the minimum iCGM specifications, virtual clinical trial testing as needed, and integration V&V testing.",
       "mark": [
        111,
        133
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K253585",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "METICULY Patient-specific titanium mesh implant",
     "meta": [
      "Meticuly",
      "14 January 2026",
      "K252958",
      "Neurology"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Table 4: Performance testing summary table Test Test method summary Results Comparative device modeling Computational simulation was The FEA consideration is based on with Finite Element Analysis performed with a compression load maximum equivalent stress, safety (FEA) in the normal direction on the factor and maximum deformation. center of the mesh implant.",
       "mark": [
        104,
        128
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K252958",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "SAPIEN M3 Transcatheter Mitral Valve Replacement System",
     "meta": [
      "Edwards Lifesciences",
      "22 December 2025",
      "P250019",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Finite element analysis To determine mechanical No fracture of SAPIEN M3 strain during dock loading, dock structural components deployment and cyclic predicted within a minimum of loading.",
       "mark": [
        0,
        14
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P250019",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "DBLG2",
     "meta": [
      "Diabeloop",
      "19 December 2025",
      "K251152",
      "Clinical Chemistry"
     ],
     "tags": [
      {
       "label": "In silico or virtual patients"
      }
     ],
     "quotes": [
      {
       "label": "FDA's decision summary",
       "text": "Once a compatible candidate iCGM is identified, the sponsor will conduct virtual clinical trial testing with virtual patients, with pre-defined acceptance criteria based on virtual patient equivalency testing and glycemic outcomes from the iAGC clinical study.",
       "mark": [
        73,
        95
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K251152",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "MiniMed 780G System",
     "meta": [
      "Medtronic Minimed",
      "11 December 2025",
      "P160017S125",
      "Clinical Chemistry"
     ],
     "tags": [
      {
       "label": "In silico or virtual patients"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Simulation studies were conducted with various setpoint (100 and 120 mg/dL) and active insulin time (AIT; 2 or 4 hours) combinations to enable the analysis of in-silico outcomes with various parameter settings.",
       "mark": [
        159,
        168
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P160017S125",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "ProVee System for BPH",
     "meta": [
      "Proverum",
      "9 December 2025",
      "P250011",
      "Gastroenterology, Urology"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "To evaluate the durability and integrity of the Expander using Finite Element Analysis (FEA) in Implant Strain simulated physiological Analysis/Fatigue Strain conditions to calculate a fatigue Fatigue Safety Factor > 1 Pass Limit and Factor of factor of safety for the Expander Safety per ASTM E739 and FDA recognized standard ASTM F3211.",
       "mark": [
        63,
        77
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P250011",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Omnipod 5 algorithm",
     "meta": [
      "Insulet",
      "3 December 2025",
      "K251779",
      "Clinical Chemistry"
     ],
     "tags": [
      {
       "label": "In silico or virtual patients"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "The changes to the software are supported by in silico clinical validation and design verification and validation which demonstrate that the subject device and predicate device are substant ially equivalent.",
       "mark": [
        45,
        54
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K251779",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "ZSFab Lumbar Interbody System",
     "meta": [
      "Zsfab",
      "25 November 2025",
      "K252610",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Additional PERFORMANCE D ATA The subject ZSFab Lumbar Interbody System was evaluated through finite element under ASTM F2077 axial compression and compression shear loading conditions and tested in dynamic axial compression and compression shear to verify that the mechanical testing under of the ZSFab Lumbar Interbody System under K242734 was the worst-case compared to subject implants.",
       "mark": [
        93,
        107
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K252610",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "BEE HA",
     "meta": [
      "NGMedical",
      "21 November 2025",
      "K250560",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Conclusion: Based on the finite element analysis and the comparison to the predicate devices, the subject device is determined to be substantially equivalent to the predicate devices.",
       "mark": [
        25,
        39
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K250560",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Dexcom Smart Basal",
     "meta": [
      "Dexcom",
      "18 November 2025",
      "K252818",
      "Clinical Chemistry"
     ],
     "tags": [
      {
       "label": "In silico or virtual patients"
      }
     ],
     "quotes": [
      {
       "label": "FDA's decision summary",
       "text": "The changes were tested in an in silico clinical trial re-run validation test to assess how changes made to the algorithm used in the clinical study impact the recommended doses compared to the recommended doses in the clinical study algorithm.",
       "mark": [
        30,
        39
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K252818",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "KLS Martin IPS Forearm System",
     "meta": [
      "KLS-Martin",
      "14 November 2025",
      "K250865",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "MRI RF and gradient heating simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Fractional wsbSAR or head SAR for an hour-long scanning session while maintaining a temperature rise of below 6 °C were determined for the worst-case devices within the Duke virtual human anatomy to align with the most recent FDA guidance document for testing and labeling of medical devices for safety in the magnetic resonance environment.",
       "mark": [
        174,
        187
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K250865",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "MiniMed 780G System",
     "meta": [
      "Medtronic Minimed",
      "29 August 2025",
      "P160017S124",
      "Clinical Chemistry"
     ],
     "tags": [
      {
       "label": "In silico or virtual patients"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "In this submission only simulation study data was presented using Medtronic’s Virtual Patient Model to support of the performance of AHCL version 5.3 with G4S even though this approach has not yet been validated.",
       "mark": [
        78,
        93
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P160017S124",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "SmartGuard technology; Predictive Low Glucose technology",
     "meta": [
      "Medtronic Minimed",
      "29 August 2025",
      "K251217",
      "Clinical Chemistry"
     ],
     "tags": [
      {
       "label": "In silico or virtual patients"
      },
      {
       "label": "Other computational model or simulation"
      }
     ],
     "quotes": [
      {
       "label": "FDA's decision summary",
       "text": "For iCGMs, re- qualification requires demonstrated continued compatibility to the minimum iCGM specifications, virtual clinical trial testing as needed, and integration V&V testing.",
       "mark": [
        111,
        133
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K251217",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Distal Xtremities System",
     "meta": [
      "TriMed",
      "20 August 2025",
      "K252061",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "FEA analysis was used to show that design differences (number of screw holes, plate thickness, plate lengths) and material differences did not have a negative impact on mechanical strength of the DXS plates.",
       "mark": [
        0,
        3
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K252061",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Arthrex iBalance Partial Knee System",
     "meta": [
      "Arthrex",
      "1 July 2025",
      "K251453",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "MRI RF and gradient heating simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Performance Data Non-clinical testing and in-vivo electromagnetic simulation demonstrated that the Arthrex iBalance Partial Knee System is “MR Conditional”.",
       "mark": [
        50,
        76
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K251453",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "CGuard Prime Carotid Stent System",
     "meta": [
      "InspireMD",
      "23 June 2025",
      "P240029",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "N/A Characterization Analysis (FEA) and strains with the device to support only fatigue analysis.",
       "mark": [
        31,
        34
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P240029",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "SI-TECHNOLOGY® SI-DESIS® X™ Sacroiliac Joint Fusion System",
     "meta": [
      "Si-Technology",
      "6 June 2025",
      "K251525",
      "Orthopedic"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "The following adoptions / justifications were performed: Performance • Steam Sterilization Adoption Testing • Static Shear Finite Element Analysis (FEA) • Cross Sectional Area Assessment • Pushout Testing Worst Case Assessment and Adoption • Usability Worst Case Assessment and Adoption The results of these studies show the subject device is substantially equivalent to the predicate devices.",
       "mark": [
        123,
        137
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K251525",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Zenith® Iliac Branch",
     "meta": [
      "Cook",
      "30 May 2025",
      "P020018S064",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Evaluate longitudinal and Analyses must result in a fatigue pulsatile fatigue performance of safety factor > 1.0 Fatigue and ZBIS and iCAST stents for PASS Durability (FEA) 400 million cycles (equivalent to 10 years).",
       "mark": [
        168,
        171
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P020018S064",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "Arthrex Spine Compression FT Screw",
     "meta": [
      "Arthrex",
      "22 May 2025",
      "K250920",
      "Unknown"
     ],
     "tags": [
      {
       "label": "MRI RF and gradient heating simulation"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "In addition, Arthrex has performed in vivo electromagnetic simulation to further establish the safety and compatibility of the system in MR environment.",
       "mark": [
        43,
        69
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K250920",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "The Karma® Fixation System",
     "meta": [
      "Spinal Elements",
      "21 May 2025",
      "K242527",
      "Unknown"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "510(k) summary",
       "text": "Biomechanical cadaveric, Finite Element Analysis, cadaveric abrasion testing were also performed.",
       "mark": [
        25,
        39
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?id=K242527",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "EndoForce Connector for Endovascular Venous Anastomosis",
     "meta": [
      "Phraxis",
      "15 May 2025",
      "P240004",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "Finite element analysis"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "Finite Element Analysis (FEA) was used to analyze the Connector under All resulting safety Finite Element Analysis simulated in-vivo loading factors must be greater PASS conditions that included than 1. radial pulsatile and cyclic axial loads.",
       "mark": [
        0,
        14
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P240004",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "OmniaSecure™ MRI SureScan™ Lead Model 3930M",
     "meta": [
      "Medtronic",
      "22 April 2025",
      "P240036",
      "Cardiovascular"
     ],
     "tags": [
      {
       "label": "In silico or virtual patients"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "It can be seen that there is a high fracture-free rate for the Model 3930M lead based on analysis of the clinical data only as well as of the clinical and virtual patients data.",
       "mark": [
        155,
        170
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P240036",
     "urlLabel": "FDA database entry"
    },
    {
     "title": "MiniMed 780G System",
     "meta": [
      "Medtronic Minimed",
      "18 April 2025",
      "P160017S118",
      "Clinical Chemistry"
     ],
     "tags": [
      {
       "label": "In silico or virtual patients"
      }
     ],
     "quotes": [
      {
       "label": "Summary of safety and effectiveness",
       "text": "The applicant presented a simulation study using their Virtual Patient Model as support of the performance of this system with Guardian 4 Sensor.",
       "mark": [
        55,
        70
       ],
       "tone": "source"
      }
     ],
     "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P160017S118",
     "urlLabel": "FDA database entry"
    }
   ],
   "note": "110 of 9,298 authorizations cite a model. 40 shown: whole sentences first, then table text, newest first in each."
  }
 ],